The article describes automation of 2-dimensional surface analysis in the aparature for laser spectroscopy (LIBS). Such analysis give us 2D map of presented chemical elements. The main part is dedicated to algorithm choosing suitable for evaluation of the image sharpness. Digital camera and ablation laser share the same focusing optics, so knowing the exact image sharpness it is possible to set sample object to focal plane. There are theoretically described and experimentaly tested different kinds of methods how to obtain a relative sharpness number: gradient based method and methods working in frequency domain. Digital noise phenomenon is also discussed. As an output a selection of suitable method has been made with respect to its speed, accuracy and durability against digital noise. and Článek popisuje řešení automatizace dvourozměrné povrchové analýzy v zařízení pro laserovou spektroskopii (LIBS), jejímž výsledkem je 2D mapa přítomnosti chemických prvků. Hlavní část je věnována výběru algoritmu pro opětovné nastavení mapovaného vzorku do ohniskové roviny objektivu, který zároveň slouží k fokusaci laserového svazku. Děje se tak pomocí analýzy ostrosti snímku vzorku. Teoreticky jsou rozebrány a experimentálně otestovány různé metody vyhodno cení ostrosti snímku, a to metoda gradientní a metody pracující s frekvenčním spektrem obrazu. Článek se zabývá také problematikou filtrace digitálního šumu. Výstupem je volba vhodného algoritmu s ohledem na rychlost, přesnost a odolnost vůči digitálnímu šumu.
Lase Induced Brakdown Spectroscopy (LIBS) and Laser Ablation Inductively Coupled Plasma Mass Spectrometry (LA-ICP-MS) were utilized for microspatian nallyses of a fossil bear (Ursus arctos) tooth dentine. The distribution of selected trace elemnts (Sr, Ba, Fe) was measured on a 26 mm x 15 mm large and 3 mm thick transversal cross section of canine tooth, The Na and Mg content together with the distribution of matrix elemnts (Ca, P) was aslo monitored within this area. It is shown that LIBS, similarly to LA-ICP-MS can be successfully utilized for fast, spatially-resolved analysis of fossi teeth samples. Rate of Sr and Ca, Sr and Ba tracers were recognized and it is possible to say, thet these elemnts are changing its concentration in according to the seasonal increments of dentine. The concetration is lower in winter, when the bear is in his hibernation, contrary to summe, which is affuent for its substances. Secondly,concentration of Sr. shows us a seasonal migration between the place of hibernating and territory where bear searched for feed. From measurement of concentrations it is possible to claim, that the bear was hunted in his season of searching for feed, when he was ensuring his fat reserfe for winter. From archaeological pint of iew, on the base of these measurements it was possible to reconstract the ehtology of the fossil brown bear, i.e the nutrition, health and migration., Miriam Nývltová Fišáková ... [et al.]., and Obsahuje seznam literatury
The first stage of the Laser Induced Breakdown Spectroscopy (LIBS) setup development for the mobile remote sensing of chemical elements is presented. In this stage the laboratory test setup has been built using Nd:YAG laser (radiation wavelength 532 nm, ''at-top'' beam profile), custom built Galilean focusing optics together with the off-axis Newtonian collection optics for plasma radiation detection. The capability of the setup has been tested in the distance of 6.2 m on selected samples with different matrices (metal, bone, paving stone, glass). The measured concentrations have been compared to results acquired by reference LA-ICP-MS setup. These test measurements have shown that the setup is capable of semiquantitative analysis at the level of mg·kg-1. and Článek popisuje první fázi stavby mobilní sestavy pro dálkovou laserovou spektroskopii (Remote LIBS) v laboratorním uspořádání na ÚFI, FSI, VUT v Brně. Sestava využívá pulzní laser o vlnové délce 532 nm s ''at-top'' profilem svazku, fokusační optiku Galileova typu a sběrnou optiku v podobě mimoosého Newtonova teleskopu. První testování aparatury proběhlo na vzdálenost 6,2 m se vzorky různých materiálů - kovové vzorky, sklo, keramický střep, dlažební kostka a lidská kost. Naměřené koncentrace vybraných prvků byly porovnány s daty získanými metodou LA-ICP-MS. Výsledky měření ukázaly, že aparatura je dostatečně citlivá pro detekci matričních (Al, Ca, P, Mg) i některých minoritních a stopových prvků (Na, Zn, Ti, Sr) a je schopna semikvantitativní analýzy v řádech mg·kg-1.