This paper shows the comparison of a vibration measurement simultaneously performed with three different methods on the same sample. The aim of the paper is to experimentally prove the capability of a newly developed method for the measurement of vibrations with amplitudes in a nanometre range. The newly developed method is based on frequency shifted time averaged digital holographic interferometry combined with phase shifting and the phase averaging principle. The methods used as a bench mark are Doppler vibrometry performed with a commercial single point vibrometer and a single point interferometer in a Michelson construction which is improved with the lock in principle. The results have been compared and very good agreements between the results are shown. and Tento článek prezentuje výsledky měření amplitud vibrací provedených současně třemi různými metodami na stejném vzorku. Cílem tohoto článku je experimentálně prokázat správnost měření nově vyvinuté metody určené pro měření amplitud vibrací s nanometrovou velikostí. Metoda je založena na frekvenčním posunu časově středované digitální holografické interferometrie v kombinaci s principy fázového posuvu a průměrováním. Jako reference slouží komerční jednobodový vibrometr založený na Dopplerově principu a jednobodový Michelsonův interferometr pracující v lock-in režimu. V článku jsou porovnány výsledky měření, které dokládají spolehlivost nově vyvinuté metody.