The purpose of this article is to summarize current knowledge of the measurement and interpretation of infrared spectra at the Spectrometric and Laser Laboratory, Meopta - optika, s.r.o., and outline a simple and easily constructed adjustment of the FTIR spectrometer for the quantitative measurement of spectral parameters of solids. There are concisely commented transmission and reflection spectra of plane-parallel plate of material with high refractive index in the MWIR and LWIR regions as well as theoretical basis used for calculating and comparing conventionally true values of the spectral properties of these spectra. The conclusion summarizes the current situation in the measurement of spectral properties of infrared lenses and objectives. and Cílem tohoto příspěvku je shrnout dosavadní poznatky o měření a interpretaci infračervených spekter ve Spektrometrické a laserové laboratoři společnosti Meopta - optika, s.r.o. a nastínit možnou jednoduchou a konstrukčně nenáročnou úpravu FTIR spektrometrů pro kvantitativní měření spektrálních parametrů pevných látek. V příspěvku jsou stručně komentována jak výsledná transmitační (resp. reflektanční) spektra testovacích planparalelních sklíček z materiálů o vysokém indexu lomu ve spektrálních oblastech MWIR a LWIR, tak i teoretická východiska použitá pro výpočet a srovnání konvenčně pravých hodnot daných spektrálních vlastností s těmito spektry. Závěr příspěvku sumarizuje současnou situaci v oblasti měření spektrálních vlastností volných infračervených čoček a objektivů.